Dwyer Series 636D Differential Pressure Transmitter

Series 636D Differential Pressure Transmitter can be used for measuring pressures of liquids, gases & vapors. All available ranges have an excellent 0.5% F.S. accuracy with a 4-20 mA Output standard or optional 1-5 VDC output. The NEMA 4 housing is an all 316 welded construction that is designed to withstand the harshest environmental conditions. With all 316L wetted materials, this transmitter is compatible with most media. These units are CSA approved explosion-proof for use in the specified hazardous locations and meet NACE standards for off-shore applications.

Product Applications

  • Monitoring Differential Pressure for: Oil & Gas Applications, Level Measurements in Non-Vented Liquid Tanks, Filtration Systems, Compressor Control, Process Control.

Service : Compatible gases, liquids or vapors.

Wetted Materials : Types 316L SS.

Accuracy : BFSL: ±0.5% F.S. (includes linearity, hysteresis & repeatability).

Stability : ±1.0% F.S./yr.

Pressure Limits : 3 x full scale differential pressure; burst: 2500 psig.

Temperature Limits : Ambient Operating: -40 to 140°F (-40 to 60°C); Process Interface: -40 to 212°F (-40 to 100°C); Storage: -40 to 212°F (-40 to 100°C).

Compensated Temperature Range : -20 to 160°F (-29 to 71°C).

Thermal Effect : ±2% F.S./50°F (reference to 77°F).

Power Requirements : 12-30 VDC for 4-20 mA outputs; 8-14 VDC for 1-5 VDC outputs, both with reverse polarity protection.

Output Signal : 4 to 20 mA DC or 1-5 VDC.

Zero and Span Adjustments : Fixed.

Response Time : 20 ms.

Loop Resistance : 900 ohms max. @ 30 VDC for current outputs. For voltage outputs, minimum lead resistance 50k ohms.

Current Consumption : 4-20 mA for current output skus; 3 mA for voltage output skus.

Electrical Connections : 2 ft, 22 AWG cable; 3/4″ female NPT conduit.

Process Connection : Two 1/2″ female NPT.

Enclosure Rating : NEMA 4 (IP65).

Mounting Orientation : ±0.05 psi/90° rotation from horizontal.

Weight : 1.8 lb (0.82 kg).

Agency Approvals : CSA approved explosion-proof for Class I, Division 1, Groups B, C and D; Class II, Groups E, F, and G; Class III.

  • NEMA 4X rated enclosure provides protection in harsh environments permitting outdoor monitoring or in areas where dust and particulate matter exists
  • Explosion-proof device for use in hazardous areas
  • Versatile, high-accuracy device for liquid or gas supports designs requiring more precise measurements in support of application

Product Literature

Manual

Catalog Pages

 

Shopping Cart