Promag H is the preferred sensor for applications with highest requirements in the food and beverage and life science industries. Combined with the Promag 55 transmitter with touch control, four line display and extended functionality like software options for solids flow, electrode cleaning or advanced diagnostics, Promag 55H offers highest accuracy in the most complex and demanding measuring tasks.

● Flexible installation concept – numerous hygienic process connections
● Energy-saving flow measurement – no pressure loss due to cross-section constriction
● Highest performance – with integrated solids measurement for demanding fluids
● Highest safety – integrated electrode cleaning
● Automatic recovery of data for servicing
● Maintenance-free – no moving parts

 

● The measuring principle is virtually independent of pressure, density, temperature and viscosity
● For demanding hygienic applications.

Device properties
● Liner made of PFA
● Sensor housing made of stainless steel (3A, EHEDG)
● Wetted materials CIP, SIP cleanable
● 4-line backlit display with touch control
● Two switch outputs
● HART, PROFIBUS PA/DP, Modbus RS485, FF, EtherNet/IP

 

 

 

  • Measuring principle

    Electromagnetic

  • Product headline

    The flowmeter for smallest flow rates with integrated solids content diagnostics. For demanding hygienic applications.

  • Sensor features

    Flexible installation concept – numerous hygienic process connections. Energy‐saving flow measurement – no pressure loss due to cross‐section constriction. Maintenance‐free – no moving parts. Liner made of PFA. Sensor housing made of stainless steel (3-A, EHEDG).

  • Transmitter features

    Highest performance – with integrated solids measurement for demanding fluids. Highest safety – integrated electrode cleaning. Automatic recovery of data for servicing. 4‐line backlit display with touch control. Two switch outputs.

  • Nominal diameter range

    DN 2…150
    1/12″…6″

  • Max. measurement error

    ±0.25%

  • Measuring range

    0…600 m3/h

  • Max. process pressure

    PN16 to 40

  • Medium temperature range

    -20…+150°C
    (-4…+302°F)

  • Degree of protection

    IP 67 (NEMA 4x)

  • Display/Operation

    Four line display with backlit
    Touch control

  • Outputs

    4…20 mA
    Pulse / frequency(active/passive)
    Relais
    Status

  • Inputs

    Status
    Current input

  • Digital communication

    HART
    PROFIBUS PA
    FOUNDATION Fieldbus

  • Hazardous area approvals

    ATEX, FM, CSA

  • Product safety

    CE, C-tick, EAC marking

  • Hygienic approvals and certificates

    3-A, EHEDG

Datasheet:  Proline Promag 55H Technical Information

 