The high modular and robust thermometer finds its use in several applications in almost all industries. An optional head transmitter, with all common communication protocols, makes the device ready to use with enhanced measurement accuracy and reliability compared to directly wired sensors. A variety of process connections, dimensions and materials offer flexible application possibilities.
● High degree of flexibility thanks to modular design with standard terminal heads as per DIN EN 50446 and customer-specific immersion lengths
● High degree of insert compatibility and design as per DIN 43772
● Fast response time with reduced/tapered tip form
● Types of protection for use in hazardous locations: Intrinsic safety (Ex i.a.), non-sparking (Ex nA)
● Types of protection for use in hazardous locations: Intrinsic safety (Ex ia), non-sparking (Ex nA)
● Head transmitter with easy selection: Analog output 4 to 20 mA, HART®, PROFIBUS® PA or FOUNDATION Fieldbus™
● Universal range of application
● Measuring range: -40 to 1100 °C (-40 to 2012 °F)
● Pressure range up to 40 bar (580 psi)
● Degree of protection: up to IP68
Characteristic / Application
metric style
modular temperature assembly
universal range of application
suitable for hazardous areas
process connection as compression fitting
without neck
incl. thermowell / protection tube (metal)
Terminal head
TA20A – aluminium – cover with screws
TA20B – plastics – screw cover
TA20D – aluminium – hinged cover high
TA20J – SS – deep-drawn w. or w/o. display
TA20R – SS – machined w. screw cover
TA20W – aluminium – cover w. snap fit
TA21E – aluminium – screw cover w. chain
TA30A – E+H, aluminium – cover low – w. or w/o. display
TA30D – E+H, aluminium – cover high
Thermowell / protection tube
welded protection tube
Insert / probe
mineral insulated (MI), flexible
Outer diameter protection tube / Insert
9,0 mm (0,35”)
11,0 mm (0,43”)
12,0 mm (0,47”)
Max. immersion length on request
up to 10.000,00 mm (393,70”)
Material protection tube/ thermowell
1.4404 (316L)
1.4571 (316Ti)
Alloy 600 (2.4816)
Alloy C276 (2.4819)
Optional coating
Not defined
Process connection
compression fitting:
G1/2″
G1″
R1/2″
R3/4″
Tip shape
straight
reduced
tapered
Surface roughness Ra
0,8 μm (31,5 μin.)
Operating temperature range
Type K:
-40 °C …1.100 °C
(-40 °F …2.012 °F)
Type J:
-40 °C …750 °C
(-40 °F …1.382 °F)
Max. process pressure (static)
at 20 °C: 40 bar (580 psi)
Accuracy
class 1 acc. to IEC 60584
Response time
depending on configuration
t50 = 6 s
t90 = 13 s
Integration head transmitter
yes (4 … 20 mA; HART; PROFIBUS PA; FOUNDATION FIELDBUS)
Ex – approvals
ATEX II
ATEX IECEx
NEPSI
IECEx
EAC ExCertification
Gost Metrology
SIL (transmitter only)